Secondary SPRT reference thermometer
-100 °C to 430 °C

3.625,00 

Model: eXacal 217 Manufacturer: Klasmeier

Description

Key features of the secondary SPRT model 217:

Temperature range: With an operating range of -100 °C to 430 °C, our secondary SPRT reference thermometer model 217 covers a mid-range temperature span for common applications. This makes it a versatile solution for a wide range of measurement requirements.

Low measurement uncertainties: The thermometer’s precision is underscored by its low measurement uncertainties of just 2.5 mK to 8.5 mK with DAkkS-accredited calibration to DIN EN ISO/IEC 17025. This high accuracy ensures reliable results across all applications.

Robustness: As a robust alternative to very sensitive SPRTs with a quartz sheath, our secondary SPRT reference thermometer stands out with its metal sheath and stable sensing resistors. This makes it ideal for long-term use.

New SPRT sensor: The newly developed SPRT temperature sensor from Klasmeier has been improved in the following areas:

  • fast response time
  • low hysteresis
  • low self-heating
  • excellent reproducibility and stability

Temperature Sensors from our Own Manufacturing Facility

Every temperature sensor of the eXacal 230 Standard Platinum Resistance Thermometer (SPRT) is manufactured with the utmost care and precision in our in-house facility in Fulda. The development and production of our own measuring resistors guarantee the highest quality and reliability.

Why Klasmeier Manufacturing?
• Craftsmanship Precision: Our SPRTs are manufactured by experienced experts who carefully monitor every detail.
• Individual Solutions: We can address specific requirements and produce customized sensors for your application.
• Quality Control: Every sensor undergoes strict testing procedures to ensure it meets the high requirements of the International Temperature Scale ITS-90.

By manufacturing in our own facility, we have full control over the entire production process – from material selection to final accredited calibration. The result: a product that meets the highest demands in measurement technology.

DAkkS-certified calibration of secondary SPRTs

In our in-house DAkkS-accredited calibration laboratory, we offer high-precision calibration services for SPRTs. Our accreditation ensures that every calibration is performed in accordance with the internationally recognized standards of DIN EN ISO/IEC 17025. This ensures seamless traceability to national and international standards.

RangeTemperature fixed pointsMeasurement Uncertainty
-38.8344 °C – 0.01 °CMercury to water3.5 mK
0.01 °C – 29.7646 °CWater to gallium3.5 mK
0.01 °C – 156.5985 °CWater to indium6 mK
0.01 °C – 231.928 °CWater to tin8.5 mK
0.01 °C – 419.527 °CWater to zinc12 mK
The stated measurement uncertainties correspond to k=2, include an ITS-90 deviation function, self-heating and hysteresis, and are specified for the respective range.

What makes our calibration special?

Traceability: Thanks to our DAkkS accreditation, every calibration is directly linked to national and international standards. This gives you the confidence that your measurement results are recognized worldwide.

Measurement uncertainty: The accuracy of our calibration is based on a careful evaluation of all uncertainty components, including the thermometer’s reproducibility, the ITS-90 deviation function, self-heating, and sensor hysteresis.

Operating range: With the detailed information from the calibration certificate, you can use your secondary SPRT with complete confidence in the temperature range from -38.8344 °C to 419.527 °C.

Learn more about our calibration laboratory:

Calibration of Standard Platinum Resistance Thermometers

Technical Specifications


Technical Specifications Model 217
Model eXacal 217
Temperature Range -100 °C to 430 °C
Nominal Resistance Pt25 (approx. 25.5 Ohms)
Connection Type Four-Wire
Connector Blade Plug
Sheath Material Metal
Thermometer Wire Platinum
Dimensions
Handle ⌀ 20 mm x 100 mm
Sheath Diameter ⌀ 6 mm
Sheath Length 500 mm
Connection Cable 2 m
Measurement Resistor / Temperature Sensor
Length 37 mm
Type Wire-Wound Ceramic Resistor
Recommended Measurement Current 1 mA
Self-Heating 2 mA at Gallium Fixed Point
W-Value W-Gallium ≥ 1.11807 (in accordance with ITS-90 requirements)
Calibration
Fixed-Point Calibration Delivered with DAkkS calibration according to DIN EN ISO/IEC 17025 (see sample calibration certificate) or with factory calibration specifying WTP, W(Ga), and W(Zn)