Pt100 Modell N medium-temperature precision thermometer (-100 °C to 420 °C)

940,00 1.985,00 

Manufacturer: Klasmeier / eXacal Modell: Pt100N

Description

The eXacal precision thermometers are used for precise temperature measurements or as calibration standards.

Temperature range: -100 °C to 420 °C

small measurement uncertainties: 6 mK to 21 mK

robust alternative to ITS-90 standard thermometer

high stability and good calibration capability

small measurement uncertainties with DAkkS calibration certificate
The thermometer can be calibrated at temperature fixed points with small measurement uncertainties.

Nicht das richtige Produkt für Sie?

Calibration

eXacal thermometers are calibrated in our own DAkkS calibration laboratory. The DAkkS accreditation ensures that all calibrations are traceable to national and international standards.

Calibration of eXacal resistance thermometers is performed at ITS-90 fixed points. It is also possible to perform calibrations according to the comparison method in calibration baths and calibration furnaces. Measurement uncertainty is calculated from many components. The largest contributors are reproducibility, ITS-90 characteristic, self-heating and hysteresis.

Measurement uncertainties

small measurement uncertainties with DAkkS calibration certificate

Temperature fixrd points from Mercury-Triplepoint -38,8344 °C to Zinc-Meltingpoint 419,527 °C

Temperature rangeTemperature Fixed pointMeasurement uncertainty
-38,8344 °C – 0,01 °CMercury to water6 mK
0,01 °C – 29,7646 °CWater to gallium6 mK
0,01 °C – 156,5985 °CWater to indium10 mK
0,01 °C – 231,928 °CWater to tin11 mK
0,01 °C – 419,527 °CWater to zinc21 mK
The specified measurement uncertainties are k=2, include an ITS-90 characteristic curve and are given for the respective range.

Technical data

Model eXacal Pt100N
Temperature range -100 °C to 420 °C
Nominal resistance Pt100 (100 Ohm)
typical annual drift 2,5 mk
Connection type Four-wire
Connecting cable temperature resistant – very low EMF stranded wires from own production
Connection banana plugs or depending on configuration
Sheath material Metal
Thermometer wire silver
Dimensions
Handle ∅ 20 mm x 100 mm
diameter 6 mm
Length 400 mm
Connection cable 2 m
 Detector
R0 100 Ohm ± 0,05 Ohm
Alpha 0,003850 ± 0,000005
Standard IEC 60751:2008
tolerance 1/10 DIN
length 25 mm
recommended measuring current 1 mA
Self-heating 5 mK to 10 mK
DAkkS Calibration
Fixed point calibration 6 mK to 21 mK
Comparison calibration 15 mK to 35 mK