Description
DAkkS (17025) Calibration of an ITS-90 temperature fixed point cell (metal or quartz sheath) by comparison with a reference fixed point cell using highly stable standard resistance thermometers.
The following measurement uncertainties are achieved:
Fixed point | Temperature | Measurement uncertainties |
Mercury | -38,8344 °C | 2,5 mK |
Water | 0,01 °C | 0,35 mK |
Gallium | 29,7646 °C | 1,5 mK |
Indium | 156,5985 °C | 4 mK |
Tin | 231,928 °C | 5 mK |
Zinc | 419,527 °C | 8,5 mK |
Aluminium | 660,323 °C | 20 mK |
Calibration time approx. 5 working days or by arrangement.