Secondary SPRT Reference Thermometer
-200 °C to 250 °C

3.100,00 €

  Model: eXacal 216 Manufacturer: Klasmeier

Description

Low-Temperature Thermometer: The robust secondary SPRT reference thermometer for accurate temperature measurements

Low-Temperature Thermometer: The robust secondary SPRT reference thermometer for accurate temperature measurements

Our low-temperature SPRT (Standard Platinum Resistance Thermometer) serves as a robust calibration standard for the most demanding applications in precision temperature measurement. Developed to meet the needs of research, industry, and quality assurance, these thermometers offer excellent accuracy and reliability. Due to their robustness, they are an attractive alternative in daily measurement work to the often very sensitive thermometers with quartz protective tubes.

Not the right product for you?

Special Features of Secondary SPRT Model 216:

Inert gas filling: The handle of the SPRT is gas-tight, allowing the thermometer to be filled with inert gas. This design protects the thermometer from environmental influences and thus improves its performance, especially at extremely low temperatures. The inert gas filling enables the use of the thermometer at temperatures down to -200 °C, which allows calibration down to the argon triple point at approx. -189 °C.

Temperature range: With an operating range of -200 °C to 250 °C, our secondary SPRT reference thermometer model 216 covers a low temperature range. This makes it a versatile solution for a wide variety of measurement requirements.

Small measurement uncertainties: The precision of the thermometer is underlined by its low measurement uncertainties of only 4.5 mK to 7 mK for a DAkkS-accredited calibration according to DIN EN ISO/IEC 17025. This high level of accuracy ensures reliable results in all areas of application.

Robustness: As a robust alternative to very sensitive SPRTs with quartz sheaths, our secondary SPRT reference thermometer is characterized by its metal sheath and stable measuring resistors. It is therefore ideal for long-term use.

New SPRT sensor: The SPRT temperature sensor newly developed by Klasmeier has been improved in the following areas:

  • fast response time
  • low hysteresis
  • low self-heating
  • excellent reproducibility and stability

DAkkS-certified calibration of low-temperature SPRTs

In our in-house DAkkS-accredited calibration laboratory, we offer high-precision calibration services for low-temperature SPRTs. Our accreditation guarantees that every calibration is performed according to the internationally recognized standards of DIN EN ISO/IEC 17025. This ensures seamless traceability to national and international standards.

Our secondary SPRT reference thermometer for low temperatures is specifically designed to be calibrated at temperatures down to the argon triple point. This enables applications in the very low temperature range that require reliable temperature measurement.

Measurement uncertainty: Secondary SPRT reference thermometer model 216
RangeTemperature fixed pointsMeasurement uncertainty
-189.3442 °C – 0.01 °CArgon to water6 mK
-38.8344 °C – 0.01 °CMercury to water6 mK
0.01 °C – 29.7646 °CWater to gallium6 mK
0.01 °C – 156.5985 °CWater to indium6 mK
0.01 °C – 231.928 °CWater to tin7 mK
The specified measurement uncertainties correspond to k=2, include an ITS-90 deviation function, self-heating, and hysteresis, and are specified for the respective range.

What makes our calibration special?

Traceability: Through our DAkkS accreditation, every calibration is directly linked to national and international standards. This gives you the assurance that your measurement results are recognized worldwide.

Measurement uncertainty: The accuracy of our calibration is based on a careful evaluation of all uncertainty components, including the reproducibility of the thermometer, the ITS-90 deviation function, the self-heating, and the hysteresis of the sensor.

Application range: With the detailed information from the calibration certificate, you can use your low-temperature SPRT in the temperature range from -189 °C to 232 °C with full confidence.

Technical Data

Technical data model 216
Model eXacal 216
Temperature range -200 °C to 250 °C
Nominal resistance Pt25 (25.5 ohms, tolerance ±1 ohm)
Connection type Four-wire
Connection Banana plugs
Sheath material Metal
Thermometer wire Platinum
Dimensions
Handle ⌀ 20 mm x 100 mm
Protective tube diameter ⌀ 6 mm
Protective tube length 400 mm
Protective tube filling Inert gas
Connection cable 2 m
Measuring resistor / Temperature sensor
Length 37 mm
Recommended measuring current 1 mA
Self-heating 2 mA at the gallium fixed point
W-value W-gallium ≥ 1.11807 (according to ITS-90 requirements)
Calibration
Fixed-point calibration

We optionally offer accredited fixed-point calibration according to DIN EN ISO/IEC 17025 within the scope of our DAkkS accreditation. Calibration is performed at ITS-90 temperature fixed points.
The temperature range and calibration points can be coordinated depending on the application and desired scope of calibration. The achievable measurement uncertainties result from the selected procedure and are documented in the calibration certificate.

Calibration certificate

With the calibration, you receive a calibration certificate that transparently documents the execution and results of the calibration. The calibration certificate contains, among other things, information on the calibrated measuring equipment, the calibration procedure used, the reference standards used, the measurement conditions, the calibration results, and the measurement uncertainty.
When calibrating SPRTs, the calibration certificate also includes two calibration curves for different measuring currents. This allows the influence of the measuring current and the resulting self-heating on the calibration results to be transparently taken into account.
This provides you with a reliable basis for the professional evaluation of measurement results as well as for your quality and test equipment monitoring.

Downloads

Manual Model 216