Secondary SPRT Reference Thermometer
-100 °C to 670 °C

4.095,00 €

Model: eXacal 215 Manufacturer: Klasmeier

Description

Features of Secondary SPRT Model 215:

Temperature Range: With an operating range of -100 °C to 670 °C, our Secondary SPRT Reference Thermometer Model 215 covers a medium temperature range for common applications. This makes it a versatile solution for diverse measurement requirements.

Low Measurement Uncertainties: The precision of the thermometer is underscored by its low measurement uncertainties, ranging from only 2.5 mK to 13.5 mK, with a DAkkS accredited calibration according to DIN EN ISO/IEC 17025. This high accuracy ensures reliable results in all areas of application.

Robustness: As a robust alternative to highly sensitive SPRTs with quartz sheaths, our Secondary SPRT Reference Thermometer features a metal sheath and stable measuring resistors. This makes it ideal for long-term use.

New SPRT Sensor: The SPRT temperature sensor newly developed by Klasmeier has been improved in the following areas:

  • Fast response time
  • Low hysteresis
  • Low self-heating
  • Excellent reproducibility and stability

Temperature Sensors from Our Own Manufacturing Facility

Every temperature sensor of the eXacal 230 Standard Platinum Resistance Thermometer (SPRT) is manufactured with the utmost care and precision in our in-house facility in Fulda. The development and production of our own measuring resistors guarantee the highest quality and reliability.

Why Klasmeier Manufacturing?
• Craftsmanship Precision: Our SPRTs are manufactured by experienced experts who meticulously monitor every detail.
• Individual Solutions: We can address specific requirements and produce customized sensors for your application.
• Quality Control: Every sensor undergoes stringent testing procedures to ensure it meets the high requirements of the International Temperature Scale ITS-90.

By manufacturing in our own facility, we maintain full control over the entire production process – from material selection to final accredited calibration. The result: a product that meets the highest demands in measurement technology.

DAkkS-Certified Calibration of Secondary SPRTs

In our in-house DAkkS-accredited calibration laboratory, we offer high-precision calibration services for SPRTs. Our accreditation guarantees that every calibration is performed according to the internationally recognized standards of DIN EN ISO/IEC 17025. This ensures seamless traceability to national and international standards.

RangeTemperature Fixed PointsMeasurement Uncertainty
-38.8344 °C – 0.01 °CMercury to Water2.5 mK
0.01 °C – 29.7646 °CWater to Gallium3.5 mK
0.01 °C – 231.928 °CWater to Tin8.5 mK
0.01 °C – 419.527 °CWater to Zinc13.5 mK
0.01 °C – 660.323 °CWater to Aluminum13.5 mK
The stated measurement uncertainties correspond to k=2, include an ITS-90 deviation function, self-heating, and hysteresis, and are specified for the respective range.

What Makes Our Calibration Special?

Traceability: Through our DAkkS accreditation, every calibration is directly linked to national and international standards. This provides you with the assurance that your measurement results are recognized worldwide.

Measurement Uncertainty: The accuracy of our calibration is based on a careful evaluation of all uncertainty components, including the reproducibility of the thermometer, the ITS-90 deviation function, the self-heating, and the hysteresis of the sensor.

Application Range: With the detailed information from the calibration certificate, you can use your Secondary SPRT with full confidence in the temperature range of -38.8344 °C to 660.323 °C.

Learn more about our calibration laboratory:

Calibration of Standard Platinum Resistance Thermometers

Technical Data

Technical Data Model 215
Model eXacal 215
Temperature Range -100 °C to 670 °C
Nominal Resistance Pt25 (25.5 Ohms, Tolerance ±1 Ohm)
Connection Type Four-wire
Connection Lamella Plug
Sheath Material Metal
Thermometer Wire Platinum
Dimensions
Handle ⌀ 20 mm x 100 mm
Protective Tube Diameter ⌀ 6 mm
Protective Tube Length 500 mm
Connection Cable 2 m
Measuring Resistor / Temperature Sensor
Length 37 mm
Type Wire-wound Ceramic Measuring Resistor
Recommended Measuring Current 1 mA
Self-heating 2 mA at Gallium Fixed Point
W-Value W-Gallium ≥ 1.11807 (according to ITS-90 requirements)
Calibration
Fixed-Point Calibration

Optionally, we offer an accredited fixed-point calibration according to DIN EN ISO/IEC 17025 within the scope of our DAkkS accreditation. Calibration is performed at ITS-90 temperature fixed points.
Temperature range and calibration points can be coordinated depending on the application and desired calibration scope. The achievable measurement uncertainties result from the chosen procedure and are documented in the calibration certificate.

Calibration Certificate

With the calibration, you receive a calibration certificate that transparently documents the execution and results of the calibration. The calibration certificate includes information on the calibrated measuring instrument, the calibration procedure used, the reference standards used, the measurement conditions, the calibration results, and the measurement uncertainty.
For the calibration of SPRTs, the calibration certificate additionally includes two calibration curves for different measuring currents. This allows for transparent consideration of the influence of the measuring current and the resulting self-heating on the calibration results.
This provides a reliable basis for the professional evaluation of measurement results and for your quality and test equipment monitoring.

Downloads

Manual Model 215