Calibration of Dry-Block Calibrators

Reliable Measurement Values for Your Temperature Dry-Block Calibrator

Your temperature dry-block calibrators are only as reliable as their last calibration.

With our DAkkS-accredited calibration according to DIN EN ISO/IEC 17025, you receive measurement results that are traceable and fully documented.

This creates a solid foundation for your quality assurance and saves you from discussions during your next audit.

Low Measurement Uncertainties

In the accredited range, measurement uncertainties range between 0.08 K and 6.0 K, depending on the temperature range.

Fast Turnaround Times

Calibration within 5 business days for minimal downtime.

Standard-Compliant Calibration

In the accredited range according to DIN EN ISO/IEC 17025 (DAkkS) based on DKD-R 5-4
(I-CAL-GUI-013).

Price Overview

Standard package with 5 calibration points

Scope of services:

  • Accredited calibration according to DIN EN ISO/IEC 17025 (DAkkS) at 5 calibration points, in standard distribution or according to your specifications.
  • Calibration certificate with documented calibration results as well as information on measurement deviation and measurement uncertainty.
  • For ISOTECH devices, controller adjustment is possible following technical inspection and your approval.

Alternative: Individual calibration points

If you require a custom number of calibration points, we calculate on a modular basis. The costs are composed as follows:

  • Base cost: €505.00
  • Per calibration point: + €100.00

Calibration according to this model is possible starting from 3 calibration points.
The scope of services corresponds to the standard package, with the number of calibration points determined individually.

Accredited Measurement Ranges and Extensions according to Scope of Accreditation

Measurement rangeMeasurement uncertaintyComparison with
-40 °C to < 0 °C0.1 KStandard
resistance thermometers
0 °C to 100 °C80 mKStandard
resistance thermometers
> 100 °C to 130 °C0.1 KStandard
resistance thermometers
> 130 °C to 200 °C0.3 KStandard
resistance thermometers
> 200 °C to 400 °C0.5 KStandard
resistance thermometers
> 400 °C to 660 °C1.0 KStandard
resistance thermometers
0 °C to 660 °C1.5 KStandard thermocouples
> 660 °C to 800 °C2.5 KStandard thermocouples
> 800 °C to 1000 °C4.0 KStandard thermocouples
> 1000 °C to 1200 °C5.0 KStandard thermocouples
> 1200 °C to 1300 °C6.0 KStandard thermocouples

Technical Requirements for Calibration according to DKD-R 5-4

For standard-compliant calibration, the technical requirements defined in DKD-R 5-4 must be met.

Immersion depth: The immersion depth of the thermometer must be at least 15 times its outer diameter.
Bore and thermal contact: The bore in the calibration insert must fit the thermometer: the inner diameter may be a maximum of 0.5 mm larger than the outer diameter of the thermometer up to 660 °C, and a maximum of 1.0 mm larger above 660 °C. Alternatively, an equally good or better thermal contact can be established using suitable heat transfer media.
Measurement zone: The temperature dry-block calibrator must have a specified measurement zone of at least 40 mm in length with sufficient temperature homogeneity.
Please contact us. We will check your device and find a suitable solution.

Finding the right solution for you together

Speak directly with our expert!


Your contact person for calibration services
and proficiency testing:

Boris Kalb

The Calibration Process: Transparent, Structured, Standard-Compliant

Calibration in our laboratory takes 5 business days.

  • Step 1: Goods Receipt and Device Acceptance
    As soon as your device arrives, we unpack it and check it for visible defects or transport damage. If we notice anything, we will contact you before taking further steps. We then create your dry-block calibrator in our system under a calibration number.
  • Step 2: Preparation in the Calibration Laboratory
    Before starting the calibration, we check the conditions in the laboratory. Then we set up the dry-block calibrator, connect it, and insert the reference thermometer into the block.
  • Step 3: Performing the Calibration
    We connect the reference thermometer and record the measurement values – either via software or manually, depending on the device. we approach the target temperatures first in ascending and then in descending order to determine the hysteresis error.
  • Step 4: Evaluation, Measurement Uncertainty, and Calibration Certificate
    The measurement values are evaluated and the measurement uncertainty is calculated according to EA-4/02 M. We create the measurement uncertainty budget according to DKD-R 5-4 and record the results in the calibration certificate.
  • Step 5: Post-processing and Return Shipment
    After calibration, we label your device and attach a calibration mark with a QR code. You can use this code to access your calibration documents in the Klasmeier Cloud at any time. We then pack the dry-block calibrator securely and send it back to you.

Digital Archive Klasmeier Cloud:
Calibration documents accessible quickly and easily

The Klasmeier Cloud – the perfect filing system for all calibration documents. On all devices calibrated by Klasmeier, you will find a QR code that provides quick and easy access to our digital archive.

Particularly practical: In addition to all calibration documents, the contact details for the respective contact person at Klasmeier are always included. This way, even in five or more years, you can find the information you need with one click if you have questions about your calibration product.

Frequently Asked Questions about the Calibration of Dry-Block Calibrators (FAQ):

What information do you need for a quote?

For a quote, we only need your company details. Please also let us know the desired calibration points, the model designation, and the serial number of your dry-block calibrator – otherwise, we will create the quote based on our standard package.

What happens if my device does not fully meet the technical requirements?

Just contact us and we will find a suitable solution for you together!

What do I need to send in for calibration?

Please always send the dry-block calibrator together with the inserts used, the temperature probe, and the appropriate power cable.

Pack the device securely for transport. Temperature dry-block calibrators are sensitive devices and should be transported upright.

Can I bring my dry-block calibrator in person?

Yes, you are welcome to drop off your device in person during our goods receiving hours.

Goods receiving hours:

Monday to Thursday: 9:00 AM to 4:30 PM
Friday: 9:00 AM to 1:00 PM

How does the return shipment work and is it insured?

After calibration, we ship your device back via our transport insurance and invoice you for the insurance costs.

Alternatively, you can organize the return shipment and insurance yourself.

Is there a simplified calibration for smaller devices or QuickCal?

The scope of calibration is determined independently of the device size or product name.

Calibration is performed in a standard-compliant and traceable manner according to the scope of accreditation.

What happens if an adjustment, controller optimization, or repair is necessary?

As part of the calibration, we document the current state of your dry-block calibrator. If we find a deviation outside the manufacturer’s specification, we will discuss the further procedure with you.

For ISOTECH devices, this may involve an adjustment of the controller or the display system – we will also coordinate this with you in advance.

We generally only carry out repairs after your approval. Depending on the device, we will check whether we can perform the repair in-house.

Can I also receive the calibration certificate in printed form or in English?

The calibration certificate is provided digitally by default.

Upon request, we can also issue the calibration certificate in printed form.
Issuance in English is also possible. Please feel free to contact us if required.

Which standards and guidelines are relevant for the calibration?

Our DAkkS accreditation is based on DIN EN ISO/IEC 17025. The DAkkS-accredited calibration of temperature dry-block calibrators is performed according to DKD-R 5-4.

In the European context, the guideline EURAMET Calibration Guide No. 13, known as EURAMET cg-13 or I-CAL-GUI-013, is also used for the calibration of temperature dry-block calibrators. For our accredited calibrations, DKD-R 5-4 is the authoritative standard.

International recognition of accredited results is supported by recognition agreements such as EA MLA (Europe) and ILAC MRA (international).

Request calibration