Pt100 Model N Medium-Temperature Precision Thermometer
(-100 °C to 420 °C)

Manufacturer: Klasmeier / eXacal Model: Pt100N

Description

eXacal precision thermometers are used for precise temperature measurements or as calibration standards.

Temperature range: -100 °C to 420 °C

small measurement uncertainties: 6 mK to 21 mK

robust alternative to ITS-90 Standard Platinum Resistance Thermometers (SPRTs)

high stability and good calibration capability

Small measurement uncertainties with DAkkS calibration certificate

Not the Right Product for You?

DAkkS calibration at temperature fixed points

Medium-temperature precision thermometers can be calibrated in our own DAkkS calibration laboratory. DAkkS accreditation ensures that all calibrations are traceable to national and international standards.

The medium-temperature precision thermometers have been designed for excellent calibration at temperature fixed points. The measurement uncertainty is calculated from many components. The largest contributions to the measurement uncertainty budget are the reproducibility of the thermometer, ITS-90 deviation function, self-heating, and hysteresis of the temperature sensor. Using the parameters from the calibration certificate, the medium-temperature precision thermometer can be used with the respective deviation function in the temperature range from -38 °C to 420 °C.

Measurement uncertainty

small measurement uncertainties with DAkkS calibration certificate

Temperature fixed points from mercury triple point -38.8344 °C to zinc freezing point 419.527 °C

RangeTemperature fixed pointMeasurement uncertainty
-38.8344 °C – 0.01 °CMercury to water6 mK
0.01 °C – 29.7646 °CWater to gallium6 mK
0.01 °C – 156.5985 °CWater to indium10 mK
0.01 °C – 231.928 °CWater to tin11 mK
0.01 °C – 419.527 °CWater to zinc21 mK
The stated measurement uncertainties correspond to k=2, include an ITS-90 deviation function, self-heating, and hysteresis, and are specified for the respective range.

Technical Data

Model eXacal Pt100N
Temperature range -100 °C to 420 °C
Nominal resistance Pt100 (approx. 100 Ohm)
Connection type Four-wire
Connection Banana plug
Sheath material Metal
Thermometer wire Silver
Dimensions
Handle ø 20 mm x 100 mm
Sheath diameter ø 6 mm
Sheath length 400 mm
Connecting cable 2 m
Measuring resistor / Temperature sensor
Length 25 mm
Recommended measuring current 1 mA
Self-heating 5 mK to 10 mK
Calibration
Fixed-point calibration 6 mK to 21 mK