DAkkS (17025) calibration platinum/palladium thermocouple at temperature fixed points

2.220,00 

The calibration of platinum/palladium thermocouples is done at ITS-90 temperature fixed points in the range from water triple point up to silver freezing point with excellent measurement uncertainties.

Delivery time: 4 - 5 working days

Description

DAkkS (17025) calibration of a platinum/palladium thermocouple at the temperature fixed points the ITS-90

Achieveable DAkkS (17025) measurement uncertainties:

Calibration point Measurement uncertainty
Water triple point 0,01 °C 200 mK
Indium melting point 156,5985 °C 200 mK
Tin melting point 231,927 °C 200 mK
Zinc melting point 419,527 °C 200 mK
Aluminium melting point 660,323 °C 200 mK
Silver melting point 961,78 °C 200 mK

In addition, an inhomogeneity scan including graphical evaluation of the thermocouple is done.

The minimum immersion depth of the thermocouple is 480mm, the maximum outer diameter 8mm.

Calibration time approx. 5 working days or by arrangement.

 

Including Klasmeier Cloud:
Our digital archive for your calibration documents

     

From now on you will find a QR code on all devices calibrated by Klasmeier,
with which you can quickly and easily access our digital archive in order to view your
calibration results.

 